PXI-20XX/25XX 6U/Panel

PXI-2000 Series 6U Option
  • Manufacturer Adlink Technology Inc.
  • Category: DIO Isolated PXI/PXIe/cPCI
  • The PXI‑20XX/25XX 6U Option is part of ADLINK’s PXI‑2000 modular instrumentation platform, designed for high‑density digital I/O applications in laboratory, industrial, and embedded automation systems. It integrates the NuDAQ 743X series isolated digital I/O technology into a PXI 6U form factor, providing robust electrical isolation, high channel density, and full software support for multiple operating systems and development environments.
- High‑Density Digital I/O:
Up to 64 isolated digital input or output channels per board, configurable by model.
- PCI‑7432 / cPCI‑7432: 32 DI + 32 DO
- PCI‑7433 / cPCI‑7433: 64 DI
- PCI‑7434 / cPCI‑7434: 64 DO

- Electrical Isolation:
Up to 5000 Vrms channel‑to‑computer isolation (2500 Vrms for certain rear I/O variants), protecting host systems from high‑voltage transients and ground loops.

- High Output Drive Capability:
Each output channel can sink up to 500 mA, supporting inductive loads such as relays and solenoids.

- Wide Input Voltage Range:
Accepts 0–24 V DC input signals with logic thresholds of 0–1.5 V (Low) and 5–24 V (High).

- High Throughput:
Supports up to 10 kHz (0.1 ms) update rate for both input and output operations.

- Interrupt Support:
Dual interrupt sources on digital input channels 0 and 1, with external interrupt triggering capability.

- Connector Interface:
Standard 100‑pin SCSI‑II connector for all models, ensuring compatibility with ADLINK’s DB‑100 series terminal boards.

- Form Factor Options:
Available in PCI, CompactPCI (3U), and CompactPCI with Rear I/O configurations, including common‑ground and common‑power variants.

- Software Integration:
Comprehensive driver suite supporting Windows 95/98/NT/2000, DOS, and major automation environments (LabVIEW®, HP‑VEE™, DASYLab™, InTouch™, InControl™, ISaGRAF™, OPC).

- Plug‑and‑Play PCI Compliance:
Fully compliant with PCI Local Bus Specification Rev. 2.1, using the PLX PCI‑9050 controller for automatic resource assignment.

Country of origin: Taiwan R.O.C.

Customs tariff number: 8473 3020 000

Comprehensive Product Summary – ADLINK PXI‑20XX/25XX 6U Option (PXI‑2000 Series 6U Option, Article 91‑37004‑001)




### 1. Product Overview

Product Name: PXI‑20XX/25XX 6U Option
Series: PXI‑2000 Series 6U Option
Article Number: 91‑37004‑001
Manufacturer: ADLINK Technology Inc.
Associated Product Family: NuDAQ® / NuIPC® 743X Series – 64‑Channel Isolated Digital I/O Boards
Document Reference: User’s Guide, Manual Rev. 3.34 (October 22, 2002), Part No. 50‑11108‑202

The PXI‑20XX/25XX 6U Option is part of ADLINK’s PXI‑2000 modular instrumentation platform, designed for high‑density digital I/O applications in laboratory, industrial, and embedded automation systems. It integrates the NuDAQ 743X series isolated digital I/O technology into a PXI 6U form factor, providing robust electrical isolation, high channel density, and full software support for multiple operating systems and development environments.




### 2. Key Features

- High‑Density Digital I/O:
Up to 64 isolated digital input or output channels per board, configurable by model.
- PCI‑7432 / cPCI‑7432: 32 DI + 32 DO
- PCI‑7433 / cPCI‑7433: 64 DI
- PCI‑7434 / cPCI‑7434: 64 DO

- Electrical Isolation:
Up to 5000 Vrms channel‑to‑computer isolation (2500 Vrms for certain rear I/O variants), protecting host systems from high‑voltage transients and ground loops.

- High Output Drive Capability:
Each output channel can sink up to 500 mA, supporting inductive loads such as relays and solenoids.

- Wide Input Voltage Range:
Accepts 0–24 V DC input signals with logic thresholds of 0–1.5 V (Low) and 5–24 V (High).

- High Throughput:
Supports up to 10 kHz (0.1 ms) update rate for both input and output operations.

- Interrupt Support:
Dual interrupt sources on digital input channels 0 and 1, with external interrupt triggering capability.

- Connector Interface:
Standard 100‑pin SCSI‑II connector for all models, ensuring compatibility with ADLINK’s DB‑100 series terminal boards.

- Form Factor Options:
Available in PCI, CompactPCI (3U), and CompactPCI with Rear I/O configurations, including common‑ground and common‑power variants.

- Software Integration:
Comprehensive driver suite supporting Windows 95/98/NT/2000, DOS, and major automation environments (LabVIEW®, HP‑VEE™, DASYLab™, InTouch™, InControl™, ISaGRAF™, OPC).

- Plug‑and‑Play PCI Compliance:
Fully compliant with PCI Local Bus Specification Rev. 2.1, using the PLX PCI‑9050 controller for automatic resource assignment.




### 3. Technical Specifications

Optically Isolated Digital Inputs
- Models: PCI‑7432, cPCI‑7432, cPCI‑7432R/P, PCI‑7433, cPCI‑7433, cPCI‑7433R
- Channels: 32 DI (7432) / 64 DI (7433)
- Input Voltage Range: 0–24 V DC
- Logic H: 5–24 V
- Logic L: 0–1.5 V
- Input Resistance: 2.4 kΩ @ 0.5 W (7432); 2.4 kΩ @ 1 W (7433)
- Isolation Voltage: 5000 Vrms (2500 Vrms for 7433R/7432R/P)
- Throughput: 10 kHz
- Interrupt Sources: DI 0 and DI 1
- Power Consumption: +5 V @ 530 mA typ (7432); +5 V @ 500 mA typ (7433)

Optically Isolated Digital Outputs
- Models: PCI‑7432, cPCI‑7432, cPCI‑7432R/P, PCI‑7434, cPCI‑7434/P, cPCI‑7434R/P
- Channels: 32 DO (7432) / 64 DO (7434)
- Output Type: Darlington transistor array (ULN2803A or TD62783)
- Output Voltage: 5 V DC min to 35 V DC max
- Sink Current:
- 500 mA max per channel (single active)
- 500 mA @ 20 % duty for all channels active (Tamb = 50 °C)
- 2.25 W max per ULN2803A device
- Isolation Voltage: 5000 Vrms (2500 Vrms for cPCI‑7432R/P)
- Throughput: 10 kHz
- On‑Board Isolated +5 V Output: 150–200 mA
- Power Consumption: +5 V @ 530–560 mA typ

General Specifications
- Connector: 100‑pin SCSI‑II
- Operating Temperature: 0 °C – 60 °C
- Storage Temperature: −20 °C – 80 °C
- Humidity: 5 – 95 % RH, non‑condensing
- Dimensions:
- PCI‑7432/33/34: 106 mm (H) × 173 mm (L)
- cPCI‑7432/33/34: Standard 3U CompactPCI
- cPCI‑7432R/33R/34R: 3U CompactPCI with rear I/O




### 4. Functionality

The PXI‑20XX/25XX 6U Option integrates the 743X digital I/O subsystem into PXI instrumentation systems. Each channel is optically isolated, providing galvanic separation between field signals and the host computer.

Operation Principle:
- Digital Inputs: Use opto‑couplers with open‑collector transistor structures. Inputs can be configured as current‑sourcing or current‑sinking depending on the external wiring (common ground or common power).
- Digital Outputs: Employ Darlington transistor arrays to sink current from external loads. Fly‑wheel diodes protect against inductive kickback when driving coils or relays.
- Interrupts: Two independent interrupt sources (INT1, INT2) can be triggered by input transitions on channels 0 and 1.
- Software Control: All I/O operations are accessible via ADLINK’s PCIS‑DASK driver library or through high‑level environments (LabVIEW, VEE, ActiveX, OPC).

Typical Applications:
- Industrial and laboratory automation
- Event and frequency counting
- Pulse generation and timing control
- Watchdog and safety interlocks
- Relay and solenoid actuation
- Data acquisition front‑end for PLC or SCADA systems

Compatibility:
- PCI and CompactPCI backplanes (32‑bit, 33 MHz)
- PXI‑2000 6U chassis integration
- Supported OS: DOS, Windows 95/98/NT/2000
- Supported APIs: C/C++, Visual Basic, Delphi, LabVIEW®, HP‑VEE™, ISaGRAF™, InTouch™, InControl™, OPC




### 5. Components & Accessories

Included Items:
- 743X Series I/O Board (model‑specific)
- ADLINK Software CD (drivers, utilities, manuals)
- Software Installation Guide
- Rear I/O transition board (DB‑100RU) for rear‑I/O variants

Optional Accessories:
- DB‑100 series terminal boards and cables
- Shielded 100‑pin SCSI‑II cables
- DIN‑rail mounting kits for terminal boards

Connectors:
- CN1: 100‑pin SCSI‑II (front I/O)
- J2: Rear I/O interface (for cPCI‑743XR variants)




### 6. Installation & Setup

Hardware Installation:
1. Power off the host system and peripherals.
2. Discharge static electricity using a grounded wrist strap.
3. Insert the board into an available PCI or CompactPCI slot.
4. Secure the bracket and reassemble the chassis.
5. Power on the system; BIOS automatically assigns I/O and IRQ resources.

Software Installation:
- Windows Plug‑and‑Play automatically detects the card.
- Install drivers from the ADLINK CD using the “Software Installation Guide.”
- Verify operation using the provided test utilities (7432UTIL, 7433UTIL, 7434UTIL).

Jumper Configuration (PCI‑7433 only):
Eight jumpers (J1–J8) enable or disable transient voltage suppressors for groups of eight input channels each.

Cabling:
Use ADLINK’s 100‑pin SCSI‑II cables to connect to external terminal boards. Observe pin assignments for DI/DO, COM, VDD, IGND, and V5V lines as specified per model.




### 7. Performance & Capabilities

- Update Rate: 10 kHz (0.1 ms) for both input and output operations.
- Isolation: Up to 5000 Vrms channel‑to‑system.
- Output Drive: 500 mA sink per channel; 35 V max output voltage.
- Input Protection: Up to 24 V DC; transient suppression available via jumpers.
- Power Requirements: +5 V DC @ 500–560 mA typical.
- On‑board Power Output: +5 V isolated @ 150–200 mA for external sensors.
- Environmental Limits: 0–60 °C operation; −20–80 °C storage.
- Reliability: Optical isolation and transient suppression ensure long‑term stability in noisy industrial environments.

Limitations:
- 32‑bit I/O access only (8‑bit/16‑bit not supported).
- Isolation excludes external cabling.
- Rear I/O pinouts differ by manufacturer; use only ADLINK‑approved backplanes.




### 8. Standards & Compliance

- Bus Interface: PCI Local Bus Specification Rev. 2.1 compliant
- Controller: PLX PCI‑9050 bridge
- Form Factors: PCI, CompactPCI (3U), PXI 6U option
- Electrical Safety: 5000 Vrms isolation (ULN2803A/TD62783 transistor arrays)
- Environmental: Designed for 0–60 °C industrial operation
- Software Standards:
- OPC DA Server compatibility
- ActiveX / COM interface (DAQBench™)
- IEC 1131‑3 SoftPLC (ISaGRAF™) driver support
- Regulatory Compliance: CE and FCC (Class A) typical for ADLINK industrial boards




### 9. Additional Information

Warranty:
- Standard two‑year limited warranty from shipment date.
- Covers defects in materials and workmanship.
- Excludes damage from misuse, improper installation, environmental factors, or unauthorized repair.
- RMA service available via ADLINK’s global support network.

Support & Service Contacts:
- Website: [www.adlinktech.com](http://www.adlinktech.com)
- Sales & Service: service@adlinktech.com
- Technical Support: automation@adlinktech.com, nudaq@adlinktech.com
- Phone: +886‑2‑82265877 Fax: +886‑2‑82265717
- Address: 9F, No. 166, Jian Yi Road, Chungho City, Taipei 235, Taiwan

Maintenance:
- Handle boards on grounded anti‑static mats.
- Avoid exposure to high humidity or corrosive environments.
- Verify connector integrity and cable shielding periodically.

Software Updates:
- Driver and library updates available via ADLINK’s website.
- PCIS‑DASK, DAQBench™, and LabVIEW® VIs updated regularly for new OS versions.

Documentation:
- User’s Guide (Rev. 3.34)
- Software Installation Guide
- PCIS‑DASK Function Reference Manual
- LabVIEW® and HP‑VEE™ Driver Manuals




### Summary

The ADLINK PXI‑20XX/25XX 6U Option (PXI‑2000 Series 6U Option) provides a robust, high‑density digital I/O solution for PXI and CompactPCI systems. Leveraging the proven NuDAQ 743X architecture, it delivers up to 64 optically isolated channels with 5000 Vrms protection, 500 mA sink capability, and 10 kHz throughput. Its modular design supports multiple configurations—input, output, or mixed I/O—and integrates seamlessly with ADLINK’s software ecosystem, including PCIS‑DASK, LabVIEW®, HP‑VEE™, and OPC interfaces.

Engineered for industrial automation, test, and measurement applications, the PXI‑20XX/25XX 6U Option combines electrical robustness, flexible software control, and long‑term reliability, making it a versatile component in any PXI‑based data acquisition or control system.
  • DIN-100S-01 - 100pin SCSI-II termination board DIN rail w/o cable